Sensor for the measurement of specific conductivity.
For applications in surface water, potable water and cooling water.
- Four electrode design with integrated temperature sensor.
- Unaffected by fouling.
- No measuring errors due to polarization effects.
- Sterilisable
- Sensor with plug.
For the use together with Swan conductivity transmitters AMI Solicon4 or AMU Solicon4.
Specifications
Recommended conductivity range | 0.1 µS/cm … 100 mS/cm |
Accuracy | ± 1% or ± 0.2 µS/cm; whichever is greater |
Cell constant k | 0.45 cm-1 |
Temperature sensor type | Pt1000, DIN class A |
Operating conditions: | |
– max. temperature | 120°C at 6.5 bar |
– max. pressure | 12 bar at 20°C |
– sterilizable at | 120°C / 5 bar / 30min. |
Materials: | |
– Body | PPSU |
– Shaft | stainless steel SS316L |
– Current electrodes | SS 316L |
– Isolation | PEEK |
– Voltage electrodes | Platinum |
– Cover | PVDF |
Electrical connection | Plug M16 male (IP67) |
Process connection | G1″ thread |
Free space around sensor tip | 20 mm |