	Automatic, continuous measurement of total
chlorine in potable water, sanitary water, cooling water and effluent.
- Continuous online determination of disinfectants based on the DPD colorimetric method (EN ISO 7393-2; APHA 4500-Cl G).
- Measurement values: total chlorine 1, total chlorine 2, dichloramine, flow and if installed pH and temperature.
- Integrated pH measurement with temperature compensation available as option.
Specifications
Automatic, continuous measurement of total chlorine in potable water, sanitary water, cooling water and effluent.
| Disinfectant measurement | Measuring range (Standard) [ppm] | Accuracy | |
| Total chlorine (tc1 & tc2) | 0.00 | 1.00 | ± 0.01 ppm | 
| 1.00 | 3.00 | ± 0.06 ppm | |
| 3.00 | 5.00 | ± 0.2 ppm | |
| Measuring range (Extended) [ppm] | |||
| 0.00 | 10 | ± 10 % | |
| Dichloramine | Calculated difference of total chlorine 1 and 2. | ||
| Measurement time tc1 | 3 -5 sec. | 
| Measurement time tc2 | 2 min. | 
| Cycle time | 3 … 60 min. | 
| pH measuring range | 2 … 12 | 
| Temperature measuring range | -30 … 100 °C | 
| Transmitter Specifications | |
| Power Supply | 100 … 240 VAC (±10%), 50/60Hz or 10-36 VDC | 
| Power consumption | max. 35 VA | 
| Sample conditions | |
| Water consumption | min. 10 l/h | 
| Water inlet pressure | 0.15 … 2 bar | 
| Sample temperature | 5 … 50 °C | 
| Panel | |
| Dimension | 400 x 850 x 200 mm | 
| Material | PVC | 
| Weight | 9.0 kg | 
Contact us to place an order.
Downloads
| Downloads | Click Here | 