Monitor AMI-II Rescon

Monitor AMI-II Rescon

AMI-II Rescon is a complete online system for continuous measurement of specific resistivity and conductivity in ultrapure water.

The AMI-II Rescon is a process analyser developed for dependable ultrapure water monitoring in critical applications, including semiconductor manufacturing. The complete system combines the AMI-II Rescon transmitter/controller, Swansensor RC U and a stainless steel QV-HFlow L130 flow cell with digital sample flow measurement. Automatic temperature compensation and automatic range switching provide accurate, continuous data for effective real-time water quality control.

Key Features

  • Integrated flow cell and sensor provide dependable sample flow measurement.
  • Automatic temperature compensation adapts measurements to varying sample conditions.
  • Automatic range selection maintains optimal measurement performance.
  • Factory-tested and calibrated using high-purity water for reliable operation.
  • Stainless steel flow cell incorporates a needle valve and digital sample flow monitoring.

 

AMI-II Rescon Specifications

The AMI-II Rescon is a complete process analyser for continuous monitoring of conductivity and specific resistivity in ultrapure water.

The system combines the AMI-II Rescon transmitter/controller, Swansensor RC U and QV-HFlow L130 flow cell with integrated sample flow measurement. All components are assembled on a 280 × 850 mm stainless steel panel and supplied ready for installation and operation.

Measurements are automatically referenced to 25 °C. Automatic temperature compensation and measurement range selection support dependable monitoring as sample conditions change.

 

Measuring ranges

Resistivity 0.01 to 18.24 MΩ·cm
Conductivity 0.055 to 1000 µS/cm
Reference temperature Measured values compensated to 25 °C

Measurement accuracy

Resistivity ±0.5% from 0.01 to 18.24 MΩ·cm
Conductivity ±0.5% from 0.055 to 20.00 µS/cm
Conductivity ±1% from 20.00 to 1000 µS/cm

Temperature measurement

Measuring range −30 to +130 °C
Accuracy ±0.2 °C from 0 to 50 °C
Resolution 0.1 °C
Temperature compensation Non-linear for high-purity water, linear with selectable coefficient, or compensation for various chemicals

Sample conditions

Flow rate 20 to 100 L/h
Sample temperature 0 to 50 °C
Sample inlet pressure Maximum 2 bar
Sample outlet pressure Pressure-free

Power supply

AC version 100–240 VAC (±10%), 50/60 Hz (±5%)
DC version 10–36 VDC
Power consumption Maximum 35 VA

 

Contact us to place an order.