Automatic measurement for trace amounts of dissolved sodium ions in high purity water applications and steam generation
- Fully automatic three-point known addition calibration in ppb range (detection limit for sodium remains: 0.001 ppb)
 - Programmable automatic regeneration of sodium electrode
 - pH controlled alkalization reagent addition for diisopropylamine via
maintenance-free air pump 
Specifications
Automatic measurement for trace amounts of dissolved sodium ions in high purity water applications and steam generation
| Sodium measurement | Sodium glass electrode, screw cap. Calomel reference electrode, screw cap  | 
| Measuring range | 0.001 ppb – 10 ppm | 
| Accuracy | ±0.005 ppb or ±10 % of reading | 
| Reproducibility | ±0.001 ppb or ±5 % of reading | 
| Response time | 120 s (90%) | 
| Automatic- | 
  | 
|  
 Power supply  | 
 
 
  | 
| Voltage | 100-240 VAC (±10%), 50/60Hz or 10-36 VDC  | 
| Power consumption | max. 35 VA | 
|  
 Sample conditions  | 
|
| Flow rate | min. 100 ml/min. | 
| Temperature | 5 … 45 °C (41 … 113 °F) | 
| Inlet pressure | 0.3 … 3 bar (4 … 43 PSI) | 
| Outlet pressure | ambient pressure | 
| pH Value | ≥ pH 7.0 | 
| Ammonium concentration | < 10 ppm | 
| Acidity | < 50 ppm (CaCO3) | 
| Suspended solids | < 10 ppm, no oil, no grease | 
For further information see data sheet in downloads.