Monitor for simultaneous, continuous measurement of pH and redox (ORP) in potable water and effluents.
- Dual channel transmitter AMI pH/mV:pH/mV
- Flow cell M-Flow 10-3PG with temperature sensor (NT5K)
- For use with combined sensors with reference (order seperately):
-Swansensor pH Standard and ORP Standard
-Swansensor pH AY and ORP AY
Specifications
Dual channel monitor for simultaneous, continuous measurement of pH and Redox (ORP) potential in potable water and effluents.
| pH Measurement | |
| Measuring range with Swansensor ST/AY | pH 1 to 13 |
| Measuring range with Swansensor SI | pH 1 to 12 |
| Resolution | 0.01 pH |
| Reference temperature | 25° C |
| Automatic temperature compensation according to Nernst. | |
| ORP (Redox) measurement | |
| Measuring range with Swansensor ST/AY | -400 to 1’200 mV |
| Measuring range with Swansensor SI | -500 to 1’500 mV |
| Resolution | 1 mV |
| Temperature measurement | |
| Measuring range | -30 to 130°C |
| Resolution | 0.1° C |
| Power Supply | |
| Voltage | 100-240 VAC (±10%), 50/60Hz or 10-36 VDC |
| Sample conditions | |
| Flow rate | 4 to 15 l/h |
| Temperature | up to 50° C |
| Inlet pressure | up to 1 bar |
| Outlet pressure | pressure free |
| Dimensions | |
| Panel | 280 x 850 x 150 mm |
For further information see data sheet in downloads.
Contact us to place an order.
Downloads
| Downloads | Click Here |