Monitor AMI Codes-II TC

Monitor AMI Codes-II TC

Automatic, continuous measurement of total
chlorine in potable water, sanitary water, cooling water and effluent.

  • Continuous online determination of disinfectants based on the DPD colorimetric method (EN ISO 7393-2; APHA 4500-Cl G).
  • Measurement values: total chlorine 1, total chlorine 2, dichloramine, flow and if installed pH and temperature.
  • Integrated pH measurement with temperature compensation available as option.

Specifications

Automatic, continuous measurement of total chlorine in potable water, sanitary water, cooling water and effluent.

Disinfectant measurement Measuring range (Standard) [ppm] Accuracy
Total chlorine (tc1 & tc2) 0.00 1.00 ± 0.01 ppm
1.00 3.00 ± 0.06 ppm
3.00 5.00 ± 0.2   ppm
Measuring range (Extended) [ppm]
0.00 10 ± 10  %
Dichloramine Calculated difference of total chlorine 1 and 2.
Measurement time tc1 3 -5 sec.
Measurement time tc2 2 min.
Cycle time 3 … 60 min.
pH measuring range 2 … 12
Temperature measuring range -30 … 100 °C
 

Transmitter Specifications

Power Supply 100 … 240 VAC (±10%), 50/60Hz or 10-36 VDC
Power consumption max. 35 VA
 

Sample conditions

Water consumption min. 10 l/h
Water inlet pressure 0.15 … 2 bar
Sample temperature 5 … 50 °C
 

Panel

Dimension 400 x 850 x 200 mm
Material PVC
Weight 9.0 kg

 

Contact us to place an order.