Monitor for conductivity before and after cation exchange (CACE) with conventional resin columns.
- Calculation of pH and alkalizing agent concentration
- Reduced gaps in monitoring from column changes via automatic monitoring of the resin consumption
- Selectable temperature compensations (for all common alkali-zing agents and strong acids)
Specifications
Monitor for conductivity before and after cation exchange (CACE) with conventional resin columns.
Transmitter AMI Deltacon Power with two conductivity sensors UP-Con1000-SL and flow cell Catcon-Plus-SL mounted on a stainless steel panel, ready for installation and operation.
Transmitter with two signal outputs for measured values or continuous control, two relays as limit switches or as controllers, alarm relay as summary alarm indication, input for potential free contact.
The flow cell is equipped with a manual flow adjustment valve, a digital sample flow meter and an integrated cation exchanger vessel with automatic deaeration.
Pre-rinse setup with second resin vessel as option.
- Measuring range:
– 0.055 to 1000 μS/cm,
– pH 7.5 to 11.5,
– ammonia 0.01 to 10 ppm - Sample conditions:
– Temperature: up to 50 °C
– Inlet pressure: up to 2 bar
– Outlet pressure: pressure-free (atmospheric drain) - Sample connections:
– Inlet: Swagelok ¼” tube adapter
– Outlet: G ½” adapter for flexible tube 20 x 15 mm - Power Supply
-100-240 VAC (±10%)
-10-36 VDC
Please note that the cation exchange resin is not included in the delivery as it’s shelf is limited. Please order cation exchange resin separately.
For further information see data sheet in downloads.
Contact us to place an order.