Completely automated analyser for reactive silica measurement in trace concentrations for high-purity water applications.
The AMI Silitrace is a fully automated analyser for online measurement of trace silica in high-purity water and water-steam cycles. The determination of silica is done by the photometric analysis of molybdate blue at 815 nm. With integrated constant-temperature reaction chamber and reaction time regulation, a very high precision and low-level detection are achieved. Its user-friendly operation and reliable performance support optimal efficiency and protection of critical equipment in high purity water applications.
Key Features
- Detection limit as low as 0.5 ppb for ultra-sensitive silica monitoring.
- Automatic sample heating and regulated reaction time for the highest precision.
- Programmable, automatic calibration and verification for reliable trace-level measurements.
- Automatic zero measurement ensures high stability and precision.
- Reagent-saving mode reduces operating costs.
- Optional second sample channel or sequencer for up to six sample streams.
Silitrace Specifications
Silica measurement
| Measuring range | 0.5 to 1000 ppb |
|---|---|
| Reproducibility | ± 0.5 ppb or ± 5% (whichever is greater) |
| Cycle time | 3 minutes |
Sample conditions
| Flow rate | Min. 3 L/h |
|---|---|
| Temperature | 5 to 50 °C (41 to 122 °F) |
| Inlet pressure | 0.15 to 2 bar (2 to 29 PSI) |
| Outlet pressure | Ambient pressure |
Power supply
| Voltage | 100 to 240 VAC, 50/60 Hz |
|---|---|
| Power consumption | Max. 50 VA |