Monitor AMI Codes-II CC

Monitor AMI Codes-II CC

Colorimetric online analyser for determining various disinfectants, such as free, total and combined chlorine. 100-240 VAC, 50/60 Hz

The AMI Codes-II CC continuously measures the concentration of disinfectants such as free chlorine, total chlorine, and combined chlorine expressed as well as monochloramine and dichloramine. The analytical principle is based on the well-established DPD colorimetric method (EN ISO 7393-2 and APHA 4500-Cl G). Diagnostic functions like automatic zero-point determination or flow control enhance measurement accuracy and reliability. A pH measurement or a cleaning module can be added to increase operational efficiency.

Key Features

  • Simultaneous analysis of free, total and combined chlorine.
  • Adjustable measurement intervals reduce operating costs while protecting health and the environment.
  • User-friendly solid-state verification kit for the photometer to maintain measurement accuracy.
  • Self-diagnostics such as automatic zero-point determination and flow control enhance accuracy and reliability.
  • No measurement interference from seawater, effluents, or additives such as corrosion inhibitors and antiscalants.
  • Complete plug-and-play system, wet bench-tested and calibrated, ready for operation.

 

 

 

Codes-II CC Specifications

Disinfectant measurement (Measuring range [ppm] / Accuracy)

Free chlorine, monochloramine, total residual chlorine 0.00 – 1.00 ppm / ± 0.01 ppm
Free chlorine, monochloramine, total residual chlorine 1.00 – 3.00 ppm / ± 0.06 ppm
Free chlorine, monochloramine, total residual chlorine 3.00 – 5.00 ppm / ± 0.2 ppm

Combined chlorine Calculated as the difference between total residual chlorine and free chlorine
Measurement time 2 minutes
Cycle time (free chlorine) 1 to 12 minutes
Cycle time (total residual chlorine) 6 to 60 minutes
pH measuring range 2 to 12
Temperature measuring range -30 to 100 °C

Transmitter specifications

Power supply 100 – 240 VAC (±10%), 50/60 Hz or 10 – 36 VDC
Power consumption Max. 35 VA

Sample conditions

Water consumption Min. 10 L/h
Water inlet pressure 0.15 to 2 bar
Sample temperature 5 to 50 °C

Panel

Dimensions 400 × 850 × 200 mm
Material PVC
Weight 9.0 kg

 