Amperometric measurement of dissolved oxygen
- Integrated air pressure and temperature compensations for simple calibration using ambient air
- Long-term stable measuring system with robust electrode for low-priced operation
- Easy to handle membrane and electrolyte exchange
Specifications
Amperometric measurement of dissolved oxygen
Dissolved Hydrogen (H2)
0 – 800 ppm
Saturation
0 – 50 %
Dissolved oxygen measurement | with Swansensor Oxysafe 1000 including Pt1000 temperature sensor |
Measuring range | 0.01 to 20 ppm |
Resolution | 0.01 ppm |
Saturation (Resolution) | 0 to 200% (0.1%) |
Accuracy | 0.3% (if cal. temp. = measuring temp.) respectively: 1.5% at +-10° deviation to cal. temp. |
Response time | t90 < 180 sec. (rising concentration) |
Temperature measurement | (Pt1000) |
Measuring range | -30 to 130°C |
Resolution | 0.1° C |
Power Supply | |
Voltage | 100-240 VAC (±10%), 50/60Hz or 10-36 VDC |
Sample conditions | |
Flow rate | 4 to 15 l/h |
Temperature | up to 50° C |
Inlet pressure | up to 1 bar |
Outlet pressure | pressure free |
pH not lower than 4, suspended solids less than 10ppm |
|
Dimensions | |
Panel | 280 x 850 x 150 mm |
For further information see data sheet in downloads.
Contact us to place an order.
Downloads
Downloads | Click Here |