Monitor AMI Turbiwell 7027 AC

Monitor AMI Turbiwell 7027 AC

High precision, non-contact nephelometer based on ISO 7027 (EN 27027, DIN 38404) for the continuous measurement of turbidity in pure water, potable water, surface water treatment and effluent.

  • Range: 0.000 to 200.0 FNU with a precision of  ± 0.003 FNU or 1% of reading.
  • Turbidimeter with sample chamber and valve for chamber drain.
  • Optional sample degasser for the prevention of interfering bubbles formation in the sample.
  • High precision secondary standards for quality assurance.

Specifications

  • Non-contact turbidity (Nephelometer according to ISO 7027) with control electronics, sample chamber, drain value for chamber purging (manual or automated) and sample degasser (optional)
  • Complete system mounted on PVC panel (400 x 850mm), factory calibrated (formazine) and ready for operation
  • Measuring and control transmitter on an aluminium case (IP 66). Two signal outputs freely scalable and with simulation mode. Potential-free alarm contact as summary alarm indication for programmable alarm values and for instrument faults. Monitoring of maintenance requirements.

 

Measuring Range:

  • 0 – 200.0 FNU/NTU

 

Precision:

  • ± (0.003 FNU + 1% of reading)

 

Accuracy (based on formazine primary standard):

  • Range 0 – 40 FNU ± (0.01 FNU + 2% of reading)
  • Range > 40 FNU: ± 5% of reading

 

Sample Conditions:

  • Flow rate: Approx. 20 – 60 l/h

 

Temperature:

  • up to 45°C

 

Power Supply:

  • 100 – 240 VAC (±10%)
  • 10 – 36 VDC

 

Contact us to place an order.

Downloads

Downloads Click Here