Contact-free turbidity measurement; approved alternative method to US EPA 180.1 / ISO 7027
- No consumables, no wearing parts, no follow-up costs
- Heated optics prevent measurement errors and fouling from condensation
- Automatic measuring chamber flushing; trouble-free operation without manual intervention
- Fast and easy verification with primary and secondary standard
Specifications
Turbidity Meter Australia
Measuring Range:
- 0 – 200.0 NTU
Precision:
- ± (0.003 NTU+ 1% of reading)
Accuracy (based on formazine primary standard):
- Range 0 – 40 NTU± (0.01 FNU + 2% of reading)
- Range > 40 NTU: ± 5% of reading
Sample Conditions:
- Flow rate: Approx. 5 – 16 gal/h
Temperature:
- up to 113 °F
Power Supply:
- 100 – 240 VAC (±10%)
- 10 – 36 VDC
- Non-contact turbidity (Nephelometer according to EPA 180.1) with control electronics, sample chamber, drain value for chamber purging (manual or automated) and sample degasser (optional)
- Complete system mounted on PVC panel (15,75 x 33,5″), factory calibrated (formazine) and ready for operation
- Measuring and control transmitter on an aluminium case (IP 66). Two signal outputs freely scalable and with simulation mode. Potential-free alarm contact as summary alarm indication for programmable alarm values and for instrument faults. Monitoring of maintenance requirements.